Applications include isolation and indentification of plant/human pathogens, work in virology, tissue culture, sterility testing and recombinant DNA research. This Biohazard Cabinet offers operator protection from low to moderate risk biological agents (or chemical carcinogens(, while providing a sterile work environment for the procedure. The cabinet is kept under negative pressure with partial quantity of recirculated air exhausted through a HEPA filter. External exhaust from the laboratory is optional.
The Biological Safety Cabinet is similar in performance and specification to the Vertical Series, described on reserse.
The additional main features are as under:
AIR FLOW : 70% of the air is recirculated back into the work area and 30% of the air is HEPA filtered and exhausted out to the environment via a dedicated duct without mixing with the recirculated air.
Outer Body : (Wooden / Mild Steel Duly Powder Coated / S.S. 304)
WORKING SIZE: 2âà 2â 3â à 2â 4â à 2â 6â à 2â
FACE VELOCITY : A face velocity of 90±20 FPM is maintained.
Optional :